cms-4000 Continuous monitoring system
The CMS-4000 is a simple, reliable tool to measure and monitor total petroleum hydrocarbons in water. With no moving parts and corrosion protected components, the system can be utilized in many different applications.
The CMS-4000 logger with the Fiber Optic Chemical Sensor (FOCS®) detects the presence of petroleum hydrocarbons in water.
The FOCS probe takes advantage of the interaction between the light traveling through a fiber and a water solution containing petroleum hydrocarbons. As the concentration of hydrocarbons increases, the light scattered from the proprietary probe fiber increases in a quantitative relationship. The probe output is monitored by a controller that has 0-5 volt or 4-20 mA output and remote access capability. The system can monitor up to four (4) probes. The probe is intrinsically safe* and designed to be remotely mounted in hazardous locations.
*When used with the FCI Environmental Intrinsic Safety Barrier.
APPLICATIONS
- Process water
- Waste water
- Oil field produced water
- Separation vessel effluent
- Storm water run-off
- Bilge and ballast water
- Groundwater remediation monitoring
- Carbon filter bed breakthrough
- Heat exchanger leak detection
FEATURES
- Reliable tool that measures petroleum
hydrocarbons in water - Used in applications where Web connection
is NOT required - Used where logger needs to communicate with
other equipment in the plant - Continuous data collection and recording
- Intrinsically safe probe network*
- Analog (V or mA) output
- Connect up to 4 DHP hydrocarbon probes
- Real time remote monitoring
*Optional
PERFORMANCE AND HARDWARE SPECIFICATIONS OF THE CMS-4000
| CMS-4000 PERFORMANCE SPECIFICATIONS IN WATER | |
|---|---|
| Operating Range: | 0-2,000 ppm as TPH |
| Lower Limit of Detection: | 0.1 ppm as xylene |
| Hydrocarbons Detected: | C6 and higher MW petroleum hydrocarbons |
| Accuracy/Precision: | ±10% of reading |
| Response Time (initial): | <15 seconds |
| Response Time (to 95%): | <5 minutes |
| Operating Temperature Range: | 0° - 50° C |
| Trend Correlation with GC data: | 98% vs. EPA Method 8020 |
| CMS-4000 HARDWARE SPECIFICATIONS | |
|---|---|
| Enclosure: | NEMA 4X |
| Analog Output: | 0-5 Volts, 4-20 mA (optional) |
| Physical Dimensions: | 12" L X 11" W X 6.5" D. (305 X 280 X 165) |
| Enclosure Wt.: | 9 lbs. |
| Environmental Temperature: | 0°C to +50°C |
| Sample Temperature: | UL, CUL, KEMA, CE |